Skip to main content
Meloq Devices
Back to All Studies
Reliability Study2019Sweden

Measurement of Knee Joint Range of Motion with a Digital Goniometer: A Reliability Study

Authors: Svensson M, Lind V, Löfgren Harringe M

Journal: Wiley Online Library

View Original Publication

Key Finding

This study demonstrated very high intrarater and interrater reliability of the EasyAngle® digital goniometer for measuring knee joint range of motion.

Abstract

Measurements of joint range of motion (ROM) are part of a physical therapist's daily work. Activities of daily living and exercises can be complicated to perform when ROM is limited, and depending on the demands in daily living, the knee joint requires different ROM.

In sports, a few degrees in ROM may make the difference between getting injured or not. The goals for physical therapists are to help the patients to regain full ROM, mobility, strength, and function after sustaining an injury.

To measure joints with the manual universal goniometer is considered time-consuming and difficult with respect to repeated measurements. Recently, a new digital instrument for measuring range of motion was developed-EasyAngle.

The first objective of the study was to investigate the reliability of EasyAngle for measuring knee joint angles, considering intrarater and interrater reliability. The second objective was to investigate if there were any differences in the intrarater reliability between a novice and an experienced assessor.

Methods

Passive knee angles were measured in fixed positions for 40 knee joints (20 subjects). Two registered physical therapists, one novice and one experienced, conducted the measurements. Both registered physical therapists were blinded to the measurements throughout the study.

Results

The results showed very good intrarater (intraclass correlation coefficient [ICC] 0.997-0.998, standard error of mean 1.15-1.48, smallest detectable difference [SDD] 3.19-4.09, limits of agreement -3.36-3.04, -4.66-4.09) and interrater reliability (ICC 0.994, standard error of mean 2.11, SDD 5.85, limits of agreement -4.75-6.95) for measurements of knee joint ROM. No statistical difference was found between the two assessors regarding intrarater reliability.

Conclusion

The EasyAngle® digital goniometer demonstrates excellent reliability for measuring knee joint range of motion, supporting its use in clinical practice.